2025-09-30
Gas detectors are essential monitoring tools for industrial safety and environmental protection. The accuracy of their measurement data is directly related to personnel safety and the effectiveness of environmental monitoring. However, over time, instruments may experience errors due to environmental changes, aging, and other factors. Therefore, regular calibration is crucial to ensuring the reliable operation of gas detectors. Do you know what precautions are necessary for gas detector calibration? Below, our editors from Zetron Technology Electronics will explain:
Before calibration, ensure the gas detector is in good working condition. Check the instrument's exterior for damage, the display is clear, the buttons are sensitive, and the sensor is clean and free of contamination. Also, ensure the instrument is fully charged and, if necessary, charge or replace the battery in advance. Ensure the instrument is powered off to avoid unnecessary errors during the calibration process.
Standard gas is central to calibration, and its quality directly impacts the calibration results. Select standard gas with an accurate concentration and a reliable source, tailored to the target gas type of the instrument, and ensure that its concentration meets the requirements of relevant standards or regulations. Also, be sure to check the expiration date of the standard gas to avoid using expired gas, which can lead to inaccurate calibration results. The standard gas bottle valve and connectors should be leak-free to ensure a stable gas concentration entering the detector.
Calibration should be performed in a stable, clean environment. Try to choose a laboratory or indoor environment with good air quality to avoid contaminants such as dust and oil that may interfere with the calibration results. Also, maintain relatively stable ambient conditions such as temperature, humidity, air pressure, and wind speed to avoid drastic changes that could affect the calibration process. For example, excessively high or low temperatures can affect sensor performance, leading to calibration errors.
Before calibration, fully warm up the gas detector and complete the self-test procedure according to the instrument manual. This step ensures the instrument is in optimal operating condition and avoids calibration errors caused by incomplete instrument stabilization.
Properly connect the standard gas to the calibration port of the detector, ensuring the connection is secure and leak-free. According to the instrument manual, strictly control the flow rate of the standard gas introduced. Excessive or insufficient flow rates will affect the accuracy of the calibration results. 1 It is recommended to use a matching flowmeter to precisely control the gas flow rate to ensure calibration reliability.
Calibration typically consists of two parts: zero calibration and concentration calibration. For zero calibration, place the detector in pure air and press the calibration button to eliminate instrument drift and zero-point error. For concentration calibration, place the detector in standard gas and press the calibration button to align the displayed value with the standard gas concentration. During this operation, carefully observe the instrument reading to ensure stability before proceeding to the next step.
After calibration, it is recommended to perform an alarm function test. Introduce standard gas above the alarm setting value and observe whether the instrument promptly issues an audible and visual alarm at the set concentration. Record the alarm trip value to verify the accuracy and timeliness of the alarm function.
After calibration, verify that the calibration results meet the requirements. You can re-introduce the standard gas and observe whether the instrument readings are consistent with the standard gas concentration, or perform repeatability tests to ensure that the instrument produces similar results under the same conditions.
Detailed records should be kept of the calibration date, calibrator, standard gas information, pre- and post-calibration readings, environmental conditions, and other information, and a calibration record sheet should be completed. These records not only help track changes in instrument performance but also provide important reference for subsequent maintenance and management.
During calibration, it is important to avoid interfering gases in the environment that could affect the calibration results. For example, certain gases may cause cross-interference with the sensor, resulting in reading deviations. Therefore, calibration should be performed in a single-gas environment or with an instrument with strong anti-interference capabilities.
The calibration frequency of a gas detector should be determined based on factors such as the operating environment and instrument performance requirements. Generally, it is recommended to calibrate at least annually. In harsh environments or for high-precision applications, the calibration frequency can be increased, such as quarterly or monthly.
If the instrument exhibits abnormal readings or performance degradation, it should be calibrated promptly. 3. Personnel Training
Operators should receive professional training and become familiar with the calibration process and instrument usage. This will prevent calibration failures or instrument damage due to misoperation.
In summary, gas detector error calibration is a highly technical and meticulous task. From pre-calibration preparation to calibration operations, and then to post-calibration inspections and record-keeping, every step requires rigorous control. Only by adhering to these precautions can we ensure the accuracy and validity of calibration results, thereby guaranteeing the reliable operation of gas detectors in industrial production, environmental protection, and other fields.